Commands Listed in Alphabetical Order
Group
RF
Syntax
MARKER:PEAK:MAXimum <NR1>
MARKER:PEAK:MAXimum?
Related Commands
MARKER:PEAK:STATE, MARKER:PEAK:EXCURsion, MARKER:PEAK:
THReshold, MARKER:TYPe,
Arguments
<NR1> is an integer that represents the maximum number of peaks that could
have automatic markers.
Examples
MARKER:PEAK:MAXimum 3 sets the maximum number of peaks that could have
automatic markers to 3.
MARKER:PEAK:MAXimum? might return 11, indicating that the maximum number
of peaks that should have automatic markers has been set to 11.
MARKER:PEAK:STATE
This command switches on or off the automatic peak markers that are available for
frequency domain traces. There are up to 11 automatic markers. The maximum
number of markers can be set using the command MARKER:PEAK:MAXimum.
The automatic peak markers find amplitude peaks based upon user threshold
and excursion settings (set with the MARKER:PEAK:EXCURsion and
MARKER:PEAK:THReshold commands.)
Each automatic marker has a readout associated with it. These can be absolute or
delta readouts (set with the MARKER:TYPe command.) To list all of the peak
markers, use the command SEARCH:SPECTral:LIST?
Group
RF
Syntax
MARKER:PEAK:STATE {OFF|ON|0|1}
MARKER:PEAK:STATE?
Related Commands
MARKER:PEAK:MAXimum, MARKER:PEAK:EXCURsion,
MARKER:PEAK:THReshold, MARKER:TYPe
Arguments
OFF or 0 turns the automatic peak markers off.
ON or 1 t
urns the automatic peak markers on.