Commands Listed in Alphabetical Order
Group
Mask
Syntax
MASK:COUNt RESET
Related Commands
MASK:COUNt:HITS?
MASK:COUNt:SEG<x>:HITS?
MASK:COUNt:FAILURES?
Examples
MASK:COUNt RESET resets to zero the number of hits and failures for all mask
segments.
MASK:COUNt:FAILURES? (Query Only)
This query returns the number of pass/fail mask tests that have failed.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:COUNt:FAILURES?
Related Commands
MASK:COUNt:HITS?
MASK:COUNt:SEG<x>:HITS?
Returns
Integer which represents the number of pass/fail mask tests that have failed.
Examples
MASK:COUNT:FAILURES? might return :MASK:COUNT:FAILURES 3, showing
that three tests have failed.
MASK:COUNt:HITS? (Query Only)
This query returns the sum of all hits in all mask segments.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-371