Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:AUXout:FAILure {ON|OFF|<NR1>}
MASK:TESt:AUXout:FAILure?
Related Commands
MASK:TESt:STATE,
MASK:TESt:COMPLetion:CRITerion,
MASK:TESt:AUXout:COMPLetion
Arguments
ON turns on this feature, so that a signal is sent to the AUX:out port whenever a
mask test status changes to ‘failing’.
OFF turns off this feature.
<NR1>= 0 turns off this feature; any other value turns it on.
Examples
MASK:TESt:AUXout:FAILure ON will send a TTL signal to the AUX:out port
upon test failure.
MASK:TESt:AUXout:FAILure? might return 0, indicating that the instrument
has not been set to send a TTL signal upon failure.
MASK:TESt:COMPLetion:CRITerion
This command specifies the criterion to be used (either by waveform or by time)
for test completion during pass/fail mask testing.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:COMPLetion:CRITerion {WAVEform|TIMe}
MASK:TESt:COMPLetion:CRITerion?
2-386 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual