Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:DELay <NR3>
MASK:TESt:DELay?
Related Commands
MASK:TESt:STATE, MASK:TESt:TIME
Arguments
<NR3> is a floating point number that specifies the amount of time in seconds, to
delay the start of the mask pass/fail test.
Examples
MASK:TEST:DELAY 2.5 delays waveform evaluation by 2.5 seconds after the
start of pass/fail testing.
MASK:TEST:DELAY? might return 5, indicating that the instrument has been set
to wait 5 seconds after the start of a mask test before it evaluates the waveforms.
MASK:TESt:HARDCopy
This command will cause the ins
trument to generate a screen shot to the default
printer as soon as a pass/fail mask test fails, using the current instrumenthard
copy settings. See HARDCopy:ACTIVeprinter and other Hardcopy commands
for more information on accessing printer settings.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:HARDCopy {ON|OFF|<NR1>}
MASK:TESt:HARDCopy?
Related Commands
MASK:TESt:STATE,
HARDCopy:ACTIVeprinter
2-388 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual