Commands Listed in Alphabetical Order
MASK:USER:TR
IGTOSAMP
This command specifies the nominal time, in seconds, from the (leading edge)
trigger position to the pulse bit sampling position, to be used for testing with
auser-define
dcustommask.
NOTE. When you create a custom mask, t he horizontal position setting is
converted from percent to time, and sets the trigtosamp value. If you c hange
trigtosamp value a fter the mask has been displayed, the mask will be horizontally
offset, and the segment x values recalculated. Note that the segment x values are
the relative time between the trigger position and the mask point's x value.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Example
s.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:TRIGTOSAMP <NR3>
MASK:USER:TRIGTOSAMP?
Related Commands
MASK:CUSTom,
SAVe:MASK,
MASK:USER:RECOrdlength
Arguments
<NR3> is a floating point number that sets the time to the pulse bit sampling
position.
Examples
MASK:USER:TRIGTOSAMP 50.0000E-6 sets the time from the trigger position
to the pulse bit sampling position to 50 μS.
MASK:USER:TRIGTOSAMP? might return 100.0000E-6, indicating that a time
of 100 μS has been specified.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-403