Commands Listed in Alphabetical Order
Group
RF
Syntax
RF:RF_AMPlitude:VERTical:SCAle <NR3>
RF:RF_AMPlitude:VERTical:SCAle?
Arguments
<NR3> is a floating point number that specifies the vertical scale.
Examples
RF:RF_AMPlitude:VERTical:SCAle 50E-3 sets the vertical scale to
50mV/division.
RF:RF_AMPlitude:VERTical:SCAle? might return 10.0000E-3, indicating
that the vertical scale has been set to 10mV/division.
RF:RF_AVErage:COUNt? (Query Only)
This query returns the number of RF traces that have been accumulated to create
the RF Average frequency domain trace.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
Group
RF
Syntax
RF:RF_AVErage:COUNt?
Returns
This command returns an integer that ranges from 0 to the number of acquisitions
specified in the command RF:RF_AVErage:NUMAVg.
Examples
RF:RF_AVErage:COUNt? might return 15, indicating that 15 traces have been
accumulated to create the RF Average trace.
RF:RF_AVErage:NUMAVg
This command specifies the number of acquisitions to be used when creating the
RF Average frequency domain trace, which displays the average of values from
multiple acquisitions at each trace point. The range is 2 – 512, in exponential
increments.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-575