Commands Listed in Alphabetical Order
Examples
SEARCH:SEARCH
1:TRIGger:A:LOGic:INPut:RF_AMPlitude HIGH
specifies to set a mark when the waveform goes above the logic threshold specified
by SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_AMPlitude.
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:R F_AMPlitude? might
return HIGH, indicating that the search engine has been set to set a
mark when a waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_AMPlitude.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:RF_FREQuency
This comm
and specifies the logic criteria to search for when searching using
the RF Frequency vs. Time trace. This command is used together with
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency to
search for an instance when the waveform goes higher or lower than the specified
threshold. “LOW” specifies to only set a mark when the RF_FREQuency
waveform drops below the logic threshold; likewise, “HIGH” specifies to seta
mark wh
en the waveform goes above the logic threshold. “X” specifies “don’t
care”.
SEARCH<x> is the search number, which is always 1.
Conditions
MDO4000/B/C models only.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut :RF_FREQuency
{HIGH|LOW|X}
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut :RF_FREQuency?
Related Commands
TRIGger:A:LOGIc:INPut:RF
Arguments
HIGH specifies a high logic level.
LOW specifies a low logic level.
X specifies a “don't care” condition.
Examples
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:R F_FREQuency HIGH
specifies to set a mark when the waveform goes above the logic threshold specified
by SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency.
SEARCH:SEARCH1:TRIGger:A:LOGic:INPut:R F_FREQuency? might
return HIGH, indicating that the search engine has been set to set a
mark when a waveform goes above the logic threshold specified by
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:RF_FREQuency.
2-724 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals