<Phase suppression> is of the range: –100 dB to 100 dB.
H INTegrate (available on instruments with the Option 2F Advanced DSP
Math only): takes the integral of the selected waveform.
<operator> ::={+|-|*|/(not available on TDS 4XXA) }
<source> ::= { CH<x> | REF<x> }
MATH2:DEFINE "Ch1 + cH2"
adds channel 1 and channel 2, and stores the result in MATH2.
MATH1:DEFINE "INV( ref4 )"
inverts the waveform stored in reference memory location 4 storing the result in
MATH1.
MATH1:DEFINE "FFT( CH1 )"
takes an FFT on the waveform from channel 1 and stores the result in MATH1.
MATH1:DEFINE "FFT( CH1, HAMM, LINEARRMS, 20 )"
takes an FFT from channel1, using the HAMMING algorithm, with linear rms
scaling, and 20 dB phase suppression. The result is stored in MATH1.
MATH1:DEFINE?
might return "Ch2*Ref2" as the expression that defines MATH1.
MATH<x>:NUMAVg
TDS 5XXA, 6XXA, & 7XXA Only,
some models require Option 2F
Allows the user to declare at what acquisition number the averaging algorithm
will begin exponential averaging. Prior to that acquisition number, the algorithm
uses stable averaging. This is equivalent to selecting Average in the Math<x>
side menu and entering a value with the general purpose knob or the keypad.
Vertical
MATH<x>:NUMAVg <NR1>
MATH<x>:NUMAVg?
Examples
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals