Command Descriptions
2-46
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
ACQuire?
Returns current acquisition settings.
ACQuire?
Might return the following string for the current acquisition:
ACQUIRE:STOPAFTER RUNSTOP;STATE 1;MODE SAMPLE;NUMAVG 16
ACQuire:MODe
Sets or queries the oscilloscope acquisition mode. This affects all
live waveforms and is equivalent to setting the Mode option in the
Acquire me nu.
Waveforms are the displayed data point values taken from
acquisition intervals. Each acquisition interval represents a time
duration that is determined by the horizontal scale (time per
division).
The oscilloscope sampling system can operate at a rate greater than
that indicated by the horizontal scale. Therefore, an acquisition
interval can include more than one sample.
The acquisition mode, which you set using this ACQuire:MODe
command, determines how the final value of the acquisition interval
is generated from the many data samples.
Acquisition
ACQuire:MODe { SAMple | PEAKdetect | AVErage }
ACQuire:MODe?
Syntax
Returns
Examples
Group
Syntax