Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-73
Might return the following string as the current cursor settings:
CURSOR:FUNCTION VBARS;SELECT:SOURCE CH1; :CUR-
SOR:VBARS:UNITS SECONDS;POSITION1 0.0E0;POSITION2
4.48E–3;:CURSOR:HBARS:UNITS VOLTS;POSITION1 0.0E0;POSI-
TION2 5.0E–1.
CURSor:FUNCtion
Selects and displays the oscilloscope cursor type. Cursors are
attached to the waveform selected by CURSor:SELect:SOUrce. This
command is equivalent to setting the Type option in the Cursor
menu. Setting the function to anything other than OFF causes the
Cursor menu to be displayed.
NOTE. Setting the display format to XY removes the cursors. Sending
the CURSor:FUNCtion command when the display format is XY
causes the oscilloscope to generate event 221 (Settings conflict) and
leaves the display in XY format.
Cursor
CURSor:FUNCtion { HBArs | OFF | VBArs }
CURSor:FUNCtion?
HBArs specifies horizontal bar cursors that measure the vertical units
in volts, amps, divisions, or decibels.
OFF removes cursors from the display.
VBArs specifies vertical bar cursors that measure time or frequency.
Group
Syntax
Arguments