Command Descriptions
2-62
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
CALibrate:INTERNal
CALIBRATE:INTERNAL
performs an internal self-calibration (SPC).
CALibrate:INTERNal:STARt
CALibrate:INTERNal:STARt (Set Only)
Performs a signal path compensation (SPC). The calibration algorithm takes a
few minutes to complete. No other command is executed until calibration is
complete. The oscilloscope state is saved before SPC execution and restored
afterwards. Pass/fail status is saved in nonvolatile memory.
Calibration and Diagnostic
CALibrate:INTERNal:STARt
CALIBRATE:INTERNAL:START
Starts a signal path compensation (SPC).
CALibrate:INTERNal:STATUS?
CALibrate:INTERNal:STATUS? (Query Only)
Returns the SPC calibration status value saved in non-volatile memory.
Calibration and Diagnostic
CALibrate:INTERNal:STATUS?
PA SS indicates that the last calibration completed without any errors detected.
FAIL indicates that the last calibration did not complete successfully.
UNCAL indicates that the non-volatile memory has failed.
Syntax
Examples
Related Commands
Group
Syntax
Examples
Related Commands
Group
Syntax
Returns

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals