Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-61
CALIBRATE:FACTORY:START
initializes the factory calibration sequence.
CALibrate:FACtory:STATUS? (Query Only)
Returns the factory calibration status value saved in non-volatile memory.
Calibration and Diagnostic
CALibrate:FACtory:STATUS?
CALIBRATE:FACTORY:STATUS?
might return :CALIBRATE:FACTORY:STATUS PASS indicating that factory
calibration passed.
CALibrate:FACtory:STEPstatus? (Query Only)
Returns pass during oscilloscope power on without regard to actual calibration
status. Returns pass or fail for the factory calibration step that most recently
executed when factory calibration is in progress. Use this query to synchronize
programmed factory calibration steps.
Calibration and Diagnostic
CALibrate:FACtory:STEPstatus?
CALIBRATE:FACTORY:STEPSTATUS?
might return :CALIBRATE:FACTORY:STEPSTATUS FAIL if the last oscilloscope
calibration step did not pass.
CALibrate:INTERNal (Set Only)
Executes the signal path compensation (SPC) calibration process. The calibration
algorithm takes a few minutes to complete. No other command is executed until
calibration is complete. The oscilloscope state is saved before SPC execution and
restored afterwards. Pass/fail status is saved in nonvolatile memory.
Calibration and Diagnostic
Examples
Group
Syntax
Examples
Group
Syntax
Examples
Group