Command Descriptions
2-86
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
CURSor:XY:RDELta? (Query Only)
(Available with TDS3AAM)
Returns the XY cursor radius delta value for the polar readout. XY display mode,
XY cursors, and Polar readout must be enabled prior to executing this command.
Cursor
CURSor:XY:RDELta?
CURSOR:XY:RDELTA?
might return CURSOR:XY:RDELTA 6.915E0.
CURSor:XY:READOUT, DISplay:XY:MODe
CURSor:XY:READOUT
(Available with TDS3AAM)
Sets or returns the XY cursor readout type.
Cursor
CURSor:XY:READOUT { POLARCoord | PRODuct | RATio | RECTangular }
CURSor:XY:READOUT?
POLARCoord sets the xy cursor readout mode to display position values in polar
coordinates.
PRODuct sets the xy cursor readout mode to display position values as the
product of the XY cursor positions.
RATio sets the xy cursor readout mode to display position values as the ratio of
the X and Y postions.
RECTangular sets the xy cursor readout mode to display postition values in
rectangular coordinates (distinct X and Y cursor positions).
CURSOR:XY:READOUT?
might return CURSor:XY:READOUT PRODUCT.
Group
Syntax
Examples
Related Commands
Group
Syntax
Arguments
Examples