Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-87
CURSor:XY:RECTX<x>? (Query Only)
Returns the XY cursor X-axis rectangular coordinate value for the specified
cursor. XY display mode, XY cursors, and Rectangular readout must be enabled
prior to executing this command.
Cursor
CURSor:XY:RECTX<x>?
<x>, where x = 1 for the reference cursor and x = 2 for the delta cursor..
CURSOR:XY:RECTX2?
might return CURSOR:XY:RECTX2 1.72E0
CURSor:XY:READOUT, DISplay:XY:MODe
CURSor:XY:RECTY<x>? (Query Only)
Returns the XY cursor Y-axis rectangular coordinate value for the specified
cursor. XY display mode, XY cursors, and Rectangular readout must be enabled
prior to executing this command.
Cursor
CURSor:XY:RECTY<x>?
<x>, where x = 1 for the reference cursor and x = 2 for the delta cursor..
CURSor:XY:RECTY1?
might return CURSOR:XY:RECTY2 -2.76E0
CURSor:XY:READOUT, DISplay:XY:MODe
Group
Syntax
Arguments
Examples
Related Commands
Group
Syntax
Arguments
Examples
Related Commands

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals