Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-95
DATA:SOURCE REF2
specifies that reference waveform two will be transferred in the next CURVE?
query.
DATA:SOURCE?
might return CH1, indicating that channel 1 is the source for the waveform data
that is transferred using a CURVE? query.
DATa:STARt
Sets or returns the starting data point for waveform transfer. This command
allows for the transfer of partial waveforms to and from the oscilloscope.
Waveform
DATa:STARt <NR1>
DATa:STARt?
<NR1> ranges from 1 to the record length, and is the first data point that will be
transferred. Data will be transferred from <NR1> to DATa:STOP or the record
length, whichever is less. If <NR1> is greater than the record length then no data
will be transferred.
When DATa:STOP is less than DATa:STARt, the values are swapped internally
for CURVe?
DATA:START 10
specifies that the waveform transfer will begin with data point 10.
DATA:START?
might return 214 as the first waveform data point that will be transferred.
CURVe?, DATa:STOP
Examples
Group
Syntax
Arguments
Examples
Related Commands

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals