Command Descriptions
2-96
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
DATa:STOP
Sets or returns the last data point that will be transferred when using the CURVe?
query. This allows the transfer of partial waveforms to the controller.
When using the CURVe command, the oscilloscope will stop reading data when
there is no more data to read or when the specified record length has been
reached so this command will be ignored.
Waveform
DATa:STOP <NR1>
DATa:STOP?
<NR1> ranges from 1 to the record length, and is the last data point that will be
transferred. If <NR1> is greater than the record length then data will be transferred
up to the record length. If both DATa:STARt and DATa:STOP are greater than
the record length, an execution error will be reported.
If you always want to transfer complete waveforms, just set DATa:STARt to 1
and DATa:STOP to the maximum record length.
When DATa:STOP is less than DATa:STARt, the values are swapped internally
for CURVe?
DATA:STOP 9000
specifies that the waveform transfer will stop at data point 9000.
DATA:STOP?
might return 1500 as the last data point that will be transferred.
CURVe?, DATa:STARt
Group
Syntax
Arguments
Examples
Related Commands

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals