Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 163
LOCk
Sets or returns the front panel lock state. This command lets you disable all
front-panel buttons and knobs to prevent manual instrument setting changes
while remotely controlling the oscilloscope. There is no front-panel equivalent.
Miscellaneous
LOCk { ALL | NONe }
LOCk?
ALL disables all front-panel controls.
NONe enables all front-panel controls. This is equivalent to the UNLock ALL
command.
NOTE. If the oscilloscope is in the Remote With Lockout State (RWLS), the LOCk
NONe command has no effect. For more information, see the ANSI-IEEE Std.
488.1-1987 Standard Digital Interface for Programmable Instrumentation,
section 2.8.3 on RL State Descriptions.
LOCK?
returns NONe when the front-panel controls are enabled by this command.
LOCk ALL
locks the front-panel controls.
UNLock
Group
Syntax
Arguments
Examples
Related Commands