Command Descriptions
2- 228
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:REFLevel:ABSolute:LOW
Sets or returns the low reference level, and is the 0% reference level when
MEASUrement:REFLevel:METHod is set to ABSolute. This command is
equivalent to setting the Reference Levels in the MEASURE menu.
Measurement
MEASUrement:REFLevel:ABSolute:LOW <NR3>
MEASUrement:REFLevel:ABSolute:LOW?
<NR3> is the low reference level, in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:LOW?
might return :MEASUREMENT:REFLEVEL:ABSOLUTE:LOW 0.0E+0 as the low
reference level.
MEASUrement:REFLevel:ABSolute:MID
Sets or returns the mid reference level, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to ABSolute. This command is
equivalent to setting the Reference Levels in the MEASURE menu.
Measurement
MEASUrement:REFLevel:ABSolute:MID <NR3>
MEASUrement:REFLevel:ABSolute:MID?
<NR3> is the mid reference level, in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:MID 2.71
sets the mid reference level to 2.71 Volts.