Command Descriptions
2- 248
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
RECALL:SETUP FACTORY
recalls (and makes current) the front panel setup to the factory defaults.
RECALL:SETUP 1
recalls the front panel setup from setup1.
RECALL:SETUP “TEK00000.SET”
recalls the front panel setup from the file TEK00000.SET in the current directory.
DELEte:SETUp, FACtory, *RCL, *RST, *SAV, SAVe:SETUp
RECAll:WAVEform (No Query Form)
Recalls a stored waveform into a reference location. This command is similar to
selecting From File in the Recall Waveform item in the SAVE/RECALL menu.
Only files saved in the internal file format can be recalled; see the SAVE:WAVE-
FORM:FILEFORMAT command.
Save and Recall
RECAll:WAVEform <file path>,REF<x>
REF<x> is the location in internal reference memory where the waveform is
recalled to.
<file path> is a quoted string that defines the file name and path. Input the file
path using the form <drive>/<dir>/<filename>. <drive> and one or more
<dir>s are optional. If you do not specify them, the oscilloscope will recall the
waveform from the current directory. <filename> stands for a filename of up to
8 characters followed by a period (“.”) and any 3-character extension. Do not use
wild card characters.
The current directory refers to the name of a directory as returned by the
FILESystem:CWD command.
RECALL:WAVEFORM “TEK00000.ISF”,REF1
recalls the waveform stored in the file named TEK00000.ISF to reference
location 1.
Examples
Related Commands
Group
Syntax
Arguments
Examples

Table of Contents