Command Descriptions
2- 256
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
*SAV (Set Only)
(Save) stores the state of the oscilloscope into a specified internal setup location.
You can later use the *RCL command to restore the oscilloscope to this saved
state. This is equivalent to selecting the Save Current Setup menu item in the
SAVE/RECALL menu.
Save and Recall
*SAV <NR1>
<NR1> is a value in the range from 1 to 10 and specifies a location. Any settings
that have been stored previously at this location will be overwritten.
*SAV 2
saves the current settings in internal setup location 2.
FACtory, *RCL, RECAll:SETUp, SAVe:SETUp
SAVe:MASK
(Requires TDS3TMT)
Saves the User mask to the specified location on the disk or flash drive.
Save and Recall
SAVe:MASK <QString>
<QString> is the path and file name, enclosed in quotes, to which to save the
User mask data on the disk or flash drive.
SAVE:SETUP “COAX2.MSK“
saves the user mask information to the file COAX2.MSK in the current directory.
Group
Syntax
Arguments
Examples
Related Commands
Group
Syntax
Arguments
Examples

Table of Contents