TRIGger:A:LOGIc:PATtern:FUNCtion { AND | NANd | NOR | OR }
TRIGger:A:LOGIc:PATtern:FUNCtion?
AND, NANd, NOR, OR set the Boolean logic operation to apply to the logic
pattern trigger inputs.
TRIGGER:A:LOGIC:PATTERN:FUNCTION NAND
sets the pattern trigger logic function to NAND.
TRIGger:A:LOGIc:PATtern:WHEn
Sets or returns the pattern logic condition on which to trigger the oscilloscope.
Trigger
TRIGger:A:LOGIc:PATtern:WHEn { TRUe | FALSe | LESSThan | MOREThan
| EQual | NOTEQual }
TRIGger:A: LOGIc: PATtern: WHEn?
TRUe triggers the oscilloscope when the pattern becomes true.
FALSe triggers the oscilloscope when the pattern becomes false.
LESSThan triggers the oscilloscope when the input pattern is true for a time
period less than the time period specified in TRIGger:A:LOGIc:PATtern:
DELTATime.
MOREThan triggers the oscilloscope when the input pattern is true for a time
period more (greater) than the time period specified in TRIGger:A:LOGIc:PAT-
tern:DELTATime.
EQual triggers the oscilloscope when the input pattern is true for a time period
equal to the time period specified in TRIGger:A:LOGIc:PATtern:DELTATime,
within a ±5% tolerance.
NOTEQual triggers the oscilloscope when the input pattern is true for a time
period greater than or less than (not equal to) the time period specified in
TRIGger:A:LOGIc:PATtern:DELTATime, within a ±5% tolerance.