Command Descriptions
2- 288
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
TRIGGER:A:PULSE:WIDTH:POLARITY NEGATIVE
specifies that the oscilloscope only trigger on negative-polarity pulse signals.
TRIGger:A:PULse:WIDth:WH En
Sets or returns the pulse width condition on which to trigger the oscilloscope.
This is equivalent to selecting the condition in the Pulse Width Trigger When
side menu.
Trigger
TRIGger:A:PULse:WIDth:WHEn { LESSThan | MOREThan | EQual |
NOTEQual }
TRIGger:A:PULse:WIDth:WHEn?
LESSThan triggers the oscilloscope when the pulse width is true for less than the
time period specified in TRIGger:A:PULse:WIDth:WIDth.
MOREThan triggers the oscilloscope when the pulse width is true for more
(greater) than the time period specified in TRIGger:A:PULse:WIDth:WIDth.
EQual triggers the oscilloscope when the pulse width is equal to the time period
specified in TRIGger:A:PULse:WIDth:WIDth, within a ±5% tolerance.
NOTEQual triggers the oscilloscope when the pulse width is greater than or less
than (not equal to) the time period specified in TRIGger:A:PULse:WIDth:WIDth,
within a ±5% tolerance.
TRIGGER:A:PULSE:WIDTH:WHEN MORETHAN
sets the oscilloscope to trigger on pulse width signals that are more (wider) than
a specified time period.
TRIGger:A:PULse:WIDth:WIDth
Sets or returns the pulse width trigger time period value.
Trigger
TRIGger:A:PULse:WIDth:WIDth <NR3>
TRIGger:A:PULse:WIDth:WIDth?
Examples
Group
Syntax
Arguments
Examples
Group
Syntax