Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 301
TRIGger:B:LEVel
Sets the B trigger level. This command is equivalent to adjusting the front-panel
TRIGGER LEVEL knob when the B trigger state is on.
Trigger
TRIGger:B:LEVel { ECL | TTL | <NR3> }
TRIGger:B:LEVel?
ECL specifies a preset ECL level of --1.3 V.
TTL specifies a preset TTL level of 1.4 V.
<NR3> is the B trigger level, in volts.
TRIGGER:B:LEVEL?
might return 1.4, indicating that the B edge trigger is set to 1.4 V.
TRIGGER:B:LEVEL 0.5
sets the B trigger level to 0.5 V.
TRIGger:B:SETLevel
Set B trigger level to 50%.
Trigger
TRIGger:B:SETLevel
TRIGGER:B:SETLevel
sets B trigger level to 50%.