Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-59
CALibrate:FACtory:NOTIfy:HOURs
Sets or returns hours when calibration is due.
Calibration and Diagnostic
CALibrate:FACtory:NOTIfy:HOURs { INFInite | <NR1> }
CALibrate:FACtory:NOTIfy:HOURs?
<NR1> = is the number of hours, 500 to 10000, before the instrument displays a
notifier saying that calibration is due. Numbers entered are rounded to the
nearest valid number.
INFInite = disables the calibration due notifier display feature.
CALIBRATE:FACTORY:NOTIFY:HOURS 1500
sets the calibration notification time to 1500 hours.
CALIBRATE:FACTORY:NOTIFY:HOURS?
might return :CALIBRATE:FACTORY:NOTIFY:HOURS 2000 indicating that the
calibration due notifier will be displayed 2000 hours after the last calibration.
CALibrate:FACtory:NOTIfy:YEARs
Sets or returns the number of years when calibration is due.
Calibration and Diagnostic
CALibrate:FACtory:NOTIfy:YEARs { INFInite | <NR3> }
CALibrate:FACtory:NOTIfy:YEARs?
<NR3> = is the number of years, 0.5 to 5, before the instrument displays a notifier
saying that calibration is due. Numbers entered are rounded to the nearest valid
number.
INFInite = disables this feature.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals