Conclusion of case
See Chapter 8 for cautions and warnings and more details on concluding a case.
1 Tear off printout. If necessary, press the “feed” button in the operate mode to
clear printout from printer cover.
Option: Print a post-case report (from the printer options setup screen).
2 Set the monitor to “standby” mode or turn it off.
3 Remove the sensor(s) from the cable-head(s).
4 Return the cable-head(s) to their docking mechanisms on the monitor.
5 Detach the H/S Probe from the CDI H/S Cuvette and return the probe to its
docking mechanism on the monitor.
6 Wipe down the monitor and cables with cleaning agent, taking care to avoid
optical surfaces.
7 Option: Disconnect the serial and power cables from the monitor, remove the
monitor from the pole clamp tray, stow the cables, and store the monitor.