choosing system settingschapter 4 4-11
The following table explains the purpose and possible values for the settings on the calcula-
tions screen.
Constant - BE Use the + and − or and keys to select a value between 10 and
47% Hematocrit. This value is used only for the calculation of Base
Excess. The default value is 25%.
Source
Flow rate input source. Choices are None, Manual, or Pump (the
default is Manual). Used in the oxygen consumption and oxygen de-
livery calculations (which can be done only if you’re using at least two
modules).
Pump Type Denes the pump type connected to the pump interface port.
Choices are “Not Used, Terumo® System 1/CDI
TM
Module (Terumo),
Sarns
TM
8000 Communications Module (Terumo), Sarns
TM
Centrifugal
Pump (Terumo), BioConsole® 550/560 (Medtronic), S5/C5
(LivaNova/Sorin/Stöckert), HL 20/Rotaow Console (Maquet/Jostra),
NEO System (Terumo).” The default is “Not Used”
Note: If you select the (LivaNova/Sorin/Stöckert), S5/C5, the arte-
rial pump position number (1-6) is required. Enter this number in the
adjacent numerical eld by using the key to highlight, then using the
and keys to adjust.
Patient BSA Patient Body Surface Area. Possible values are 0.1 to 20.0m² (in 0.05
increments), or “Not Used.” Used to index the oxygen consumption and
oxygen delivery calculations. This value automatically resets to “Not
Used” when you turn on the monitor.
Constant -
SaO
2
Possible values are 50-100% in increments of 1%. The default is 99%.
This is the arterial oxygen saturation value for use in the oxygen con-
sumption calculation if no arterial BPM is used.
Hematocrit
Offset
Possible values are -5 to +5 % in increments of 1 % (0 is default).
Saturation
Offset
Possible values are -10 to +10 % in increments of 1 % (0 is default).
Hemoglobin
Offset
Possible values are -2 to +2 g/dL in increments of 0.1 g/dL (0
is default).
Oxygen
Delivery
Possible options are Enabled and Disabled. The default setting is En-
abled.