troubleshootingchapter 99-2
Message/Problem Meaning and Corrective Action
Detected module
conguration has
changed.
System will continue
with valid modules.
Contact Terumo
Technical Service.
This error occurs if the detected module availability
does not match the modules that were installed at
the factory. You will have the option to continue using
any or all the valid modules. Please contact your
Terumo Technical Service representative to arrange
for repair.
System initialization
failure: _______
Possible messages:
Date & time lost and
needs to be reset.
Setup preferences and
calibration lost.
Previous patient case
data has been lost.
Contact Terumo
Technical Service.
There are many different system initialization failures
that can occur. Each one has an identifying code
that should be recorded for later reporting to your
Terumo Technical Service Representative. Most
failures will allow the user to continue (with certain
provisions) by pressing the √ (OK) key. Examples
are listed to the left. Failures that are not recoverable
will have no option to continue. Please contact your
Terumo Technical Service representative to arrange
for repair.
Note:
Failure code “F00D” indicates one or more
keys stuck down during startup. Ensure that all keys
on the front panel are disengaged and then reboot.
Arterial Module
(or Venous Module
or H/S Module)
initialization
failure: __________
System will con-
tinue with remaining
module(s).
Contact Terumo
Technical Service.
This indicates a module hardware failure in startup.
The user has the option to continue using the system
excluding the failed module. This failure should be
reported to your Terumo Technical Service represen-
tative to arrange for repair.
Note:
Excessive force applied to the sensor while
attached (or attaching) to the BPM cable-head can
cause the silver thermistor cap to break. The failed
BPM will need to be repaired by Terumo Technical
Service.
Barometric pressure
measurement failure:
F050
Contact Terumo
Technical Service.
This indicates an ambient pressure measurement
outside the acceptable range. A hardware failure of
the pressure reading circuit is suspected in this case.
Please contact your Terumo Technical Service repre-
sentative to arrange for repair.