troubleshootingchapter 99-14
Message/Problem Meaning and Corrective Action
Monitor lasts much
less than 25 min-
utes on back-up
battery power.
Make sure the monitor is plugged in with power off
for a full 8 hours prior to use. The green light on rear
panel indicates power being delivered to the battery.
Has the monitor been idle (uncharged) for two months
or more prior to use? If so, battery performance may
be permanently compromised. Call your Terumo Tech-
nical Service Representative to arrange for battery
replacement.
Has the battery been used (fully discharged, then
charged) repeatedly over years of service? The bat-
tery may exceed its normal life after about 200 com-
plete cycles. In this case, call your Terumo Technical
Service Representative to arrange for battery replace-
ment.
Has the service battery icon been noticed at
startup? The battery has exceeded its manufacturers
recommended replacement date.
H/S Disconnect at
Cuvette.
Connect H/S probe to cuvette.
Note:
Connecting the H/S probe to the cuvette will
deac-
tivate the alarm.

Table of Contents