9.3.4 Ambient CO
Cigarette smoke influences the measurement by more than 50 ppm.
The breath of a smoker influences the measurement by about 5 ppm.
The probe must be in the open air (CO-free) during the zeroing phase!
Open the main menu with and click on to select
Measuring modes.
Select the measurement type CO Ambient.
Carry out the measurement
Start measurement: [Start].
The reading is displayed.
Quit measurement: [Stop].
Save measured values in the app: [Save].
9.3.5 Draught measurement
Do not measure for longer than 5 min, as a drift of the pressure sensor
means that the readings could be outside the tolerance limits.
Open the main menu with and click on to select
Measuring modes.
Select the measurement type Draught.
Performing the measurement
The flue gas probe must be outside the flue.
Start measurement: [Start].
Draught zeroing is carried out.
After zeroing, position the flue gas probe in the core current (area of
the highest flue gas temperature).