The indication of the measured flue gas temperature in line 2 helps
when positioning the probe.
The reading is displayed.
Quit measurement: [Stop].
Save measured values in the app: [Save].
9.3.6 Differential pressure
DANGER
Danger of explosion due to dangerous mixture of gases!
- Before measurement close the gas path with the sealing plug, as
described below!
- If the gas path is not correctly closed with the sealing plug during
differential pressure measurement, there is a risk of unburned gas
mixture escaping.
- Escaping gas mixture can lead to an explosive atmosphere and thus
to life-threatening situations.
- Make sure there are no leaks between the sampling point and the
measuring instrument.
- Do not smoke or use naked flames during measurement, generally
avoid ignition sources.
Do not measure for longer than 5min, as a drift of the pressure sensor
means that the readings could be outside the tolerance limits.
Open the main menu with and click on to select
Measuring modes.
Select the measurement type Differential pressure.
Close the gas path with the sealing plug. Please see the description
below.