EasyManua.ls Logo

TESTO 310 II - Technical data; Measurement ranges and resolution

TESTO 310 II
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Technical data
64
11 Technical data
11.1 Measurement ranges and resolution
Measure-
ment
parameter
Measure-
ment
range
Resolution
Accuracy
Response
time t90
@ 22 °C /
72 °F
O2
0.0 ...
21.0 Vol.%
0.1 Vol.%
±0.2 Vol.%
30 s
CO
0 ...
4000 ppm
1 ppm
±20 ppm (0 ... 400 ppm)
±5 % of m.v. (401 ...
2000 ppm)
±10 % of m.v. (2001 ...
4000 ppm)
60 s
COumg
0 ...
4000 ppm
1 ppm
±20 ppm (0...400 ppm)
±5% of m.v. (401...
2000 ppm)
±10% of m.v. (2001...4000
ppm)
60 s
Draught
-0.5 hPa
+2 hPa
0.01 hPa /
0.01 InH
2
O
± 0.02 hPa / ± 0.01 InH
2
O
or ±5 % of m.v.
-
ΔP
-10.0 ...
40.0 hPa /
-4
+16 InH
2
O
0.1 hPa /
0.1 InH
2
O
± 0.5 hPa / ± 0.2 InH
2
O
(0 … 40 hPa / 0 … +16
InH
2
O)
±1 % of m.v. (rest of range)
-
Flue gas
temperature
0 ... 400 °C
/ +32 …
+752 °F
0.1 °C /
0.1 °F
± 1 °C (0 ... 100 °C) /
± 1.8 °F (+32 ... +212 °F)
± 1.5 % of m.v. (rest of
range)
<50 s
Combustion
air
temperature
-20...100°C/
-4 212°F
0,1 °C /
0.1 °F
± 1 °C /
± 1 °F
<50 s
11.2 Other instrument data
Storage and transport
temperature
-20 +50 °C / -4 +122 °F
Operating temperature
-5 +45 °C / +23 +113 °F
Humidity application
range
0 80 %RH
Operating altitude
≤ 2000 m / 6562 ft
Protection class
IP40

Table of Contents

Related product manuals