9 Smart App
48
WARNING
Hot probe shaft! Risk of burns!
> Allow the probe shaft to cool down after a measurement,
before touching it!
> Only attach the silicone hose to the probe shaft once it has
cooled down!
7
Fit silicone hose onto the probe shaft
of the flue gas probe. The probe
shaft openings must be closed.
Carry out the measurement
The silicone hose must be clear (pressureless, no kinks).
1
Start measurement: [Start].
Pressure zeroing.
2
Connect the silicone hose to the sampling point.
3
Pressurise the system.
The reading is displayed.
4
Quit measurement: [Stop].
5
Save measured values in the app: [Save].
After the measurement
1
Open filter chamber of the flue gas probe: turn it gently anti-clockwise.
2
Remove the sealing plug from the gas path.
3
Insert the particle filter again into the gas path and check that it is
fitted securely.
4
Close filter chamber of the flue gas probe.
5
Remove silicone hose from the probe shaft.