18 | Operation Ultra Low Temperature Freezers
Alarms
The Alarm Settings screen provides the option to set the
warm and cold alarm setpoints. Press the Warm Alarm or
Cold Alarm button and the setpoints screen will be displayed
permitting parameter adjustments. For more information, refer
Controls.
Figure 30. Alarm Settings Screen
Warm Alarm: The range of the warm alarm temperature
is –40°C to within 5°C of setpoint.
Note: The warm alarm will be disabled for 12 hours from a
warm start condition.
Cold Alarm: The range is –99°C to within 5°C of
setpoint.
Extreme Ambient: This allows the user to set the
extreme ambient alarm setpoint. The range is 32°C to
40°C. Default is 37°C.
Compressor Temperature: It is an alarm setpoint for
the second stage compressor sump temperature.
The range is 70°C to 98°C. Default is 94°C.
Snooze Timeout: This sets the time to snooze the
audible alarm for an active alarm.
Display
From the Settings screen, pressing the Display button will
show the Display screen. Various display settings can be
adjusted.
Figure 31. Display Screen
Brightness: Use the slide control or the +/– buttons to
adjust the brightness level of the display.
Language: To change the display language, press the
Language button and select the desired language.
Auto Date/Time: To manually set the date and time,
turn this setting off and select the format.
Units of Measure: To change the displayed unit of
measure, press the Units of Measure button.
Date: To set the date and the date format, press the Date
button.

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available

Related product manuals