48 | Appendix A: Alarm Summary Ultra Low Temperature Freezers
Control Probe
Failure
High Red Yes Yes
Message Header: Control Probe Failure
Alarm
Message Details: Cannot display cabinet
temperature. The freezer will continue to
operate in full run mode. Contact customer
service.
Ticker Message: Control Probe Failure
Alarm active. Press bell icon for more
information.
Control Probe
Failure has been
None Yellow None None
Message Header: Control Probe Failure
Recovered
Message Details: Control Probe Failure
Alarm has been cleared. Press the yellow
triangle to clear this notification.
Ticker Message: None
Heat Exchange
Probe Failure
High Red Yes Yes
Message Header: Heat Exchanger Probe
Failure Notification
Message Details: The freezer will continue to
operate with current freezer set points, but
cabinet temperature variation will increase.
Contact
customer service.
Ticker Message: Control Probe Failure
Alarm active. Press bell icon for more
information.
Heat Exchange
Probe Failure has
been
None Yellow None None
Message Header: Heat Exchanger Probe
Failure Recovered
Message Details: Heat Exchange Probe
Failure Alarm has been cleared. Press the
yellow triangle to clear this notification.
Ticker Message: None
TC1-9 Failure None Yellow None None
Message Header: TCXX Out of Range
Notification
Message Details: Information TC's has
malfunctioned. This doesn't affect the
performance of the unit. Contact service for
further assistance.
Ticker Message: None
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details