Ultra Low Temperature Freezers Appendix A: Alarm Summary | 49
TC1-9 has been None Yellow None None
Message Header: TCXX Out of Range
Notification Recovered
Message Details: Information TCXX has
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Main to UI Lost
Communication
High Red Yes Yes
Message Header: Lost Communication
Alarm
Message Details: A communication error
has occurred within the system. Contact
customer service.
Ticker Message: Lost Communication Alarm
Active. Press bell icon for more information.
Main to UI Lost
Communication
Inactive
None Yellow None None
Message Header: Lost Communication
Alarm Recovered
Message Details: A communication error
has recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
BUS Lost
Communication
High Red Yes Yes
Message Header: Backup System
Communication Alarm
Message Details: A communication error
has occurred within the back up system.
Contact customer service.
Ticker Message: BUS Lost Communication
Alarm Active. Press bell icon for more
information.
BUS Lost
Communication
Inactive
None Yellow None None
Message Header: BUS Lost Communication
Alarm Recovered
Message Details: A BUS communication
error has recovered. Select check box and
press Acknowledge to clear this notification.
Ticker Message: None
Failure to Reach
Setpoint
None Yellow None None
Message Header: Unable to Reach Setpoint
Notification
Message Details: Door openings or product
loading may cause this notification. Allow unit
to stabilize. If condition persists, contact
customer service.
Ticker Message: None
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available

Related product manuals