50 | Appendix A: Alarm Summary Ultra Low Temperature Freezers
Compressor
Temperature
None Yellow None None
Message Header: Compressor Temperature
Notification
Message Details: A refrigeration system
temperature has exceeded the compressor
temperature setpoint of XX.X C. Contact
customer service.
Ticker Message: None
Inefficient Buck /
Boost
None Yellow None None
Message Header: Inefficient buck / boost
notification
Message Details: Freezer input voltage is
out of range. Ensure proper supply voltage is
applied. If unable to clear this notification,
contact customer service.
Ticker Message: None
Power Failure Alarm High Red Yes Yes
Message Header: Power Failure Alarm
Message Details: Unit in power failure mode.
Display operating on battery power. Check
unit plug, unit circuit breaker in the ON
position, and supply voltage.
Ticker Message: Power Failure Alarm Active.
Press bell icon for more information.
Power Failure
Inactive
None Yellow None None
Message Header: Power Failure Recovered
Message Details: Power has been restored.
Press the yellow triangle to clear this
notification.
Ticker Message: None
Clean Filter Active None Yellow None None
Message Header: Clean Filter Notification
Message Details: Recommend to clean the
filter and condenser. If unable to clear
notification, contact customer service.
Ticker Message: None
Refrigeration
System Failure
High Red Yes Yes
Message Header: System Refrigeration
Failure Alarm
Message Details: An error has occurred
within the refrigeration system. Contact
customer service.
Ticker Message: System Refrigeration
Failure Alarm. Press bell icon for more
information
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details

Other manuals for Thermo Scientific TSX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSX Series and is the answer not in the manual?

Thermo Scientific TSX Series Specifications

General IconGeneral
TypeLaboratory Refrigerator
Cooling TypeForced air cooling
Defrost TypeAutomatic
Door TypeSolid or glass door options
Controller TypeMicroprocessor
Certifications/ComplianceUL, CE
Energy ConsumptionVaries by model; refer to product specification sheet
Voltage115V
Exterior Dimensions (H x W x D)Varies by model; refer to product specification sheet
Interior Dimensions (H x W x D)Varies by model; refer to product specification sheet
WeightVaries by model; refer to product specification sheet
ShelvesAdjustable shelves
Energy EfficiencyEnergy Star certified models available

Related product manuals