50 | Appendix A: Alarm Summary Ultra Low Temperature Freezers
Compressor
Temperature
None Yellow None None
Message Header: Compressor Temperature
Notification
Message Details: A refrigeration system
temperature has exceeded the compressor
temperature setpoint of XX.X C. Contact
customer service.
Ticker Message: None
Inefficient Buck /
Boost
None Yellow None None
Message Header: Inefficient buck / boost
notification
Message Details: Freezer input voltage is
out of range. Ensure proper supply voltage is
applied. If unable to clear this notification,
contact customer service.
Ticker Message: None
Power Failure Alarm High Red Yes Yes
Message Header: Power Failure Alarm
Message Details: Unit in power failure mode.
Display operating on battery power. Check
unit plug, unit circuit breaker in the ON
position, and supply voltage.
Ticker Message: Power Failure Alarm Active.
Press bell icon for more information.
Power Failure
Inactive
None Yellow None None
Message Header: Power Failure Recovered
Message Details: Power has been restored.
Press the yellow triangle to clear this
notification.
Ticker Message: None
Clean Filter Active None Yellow None None
Message Header: Clean Filter Notification
Message Details: Recommend to clean the
filter and condenser. If unable to clear
notification, contact customer service.
Ticker Message: None
Refrigeration
System Failure
High Red Yes Yes
Message Header: System Refrigeration
Failure Alarm
Message Details: An error has occurred
within the refrigeration system. Contact
customer service.
Ticker Message: System Refrigeration
Failure Alarm. Press bell icon for more
information
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details