Ultra Low Temperature Freezers Appendix A: Alarm Summary | 51
Refrigeration
System Failure
Inactive
None Red None None
Message Header: System Refrigeration
Failure Recovered
Message Details: An error in refrigeration
system has been recovered.Select check box
and press Acknowledge to clear this
notification.
Ticker Message: None
Wrong Power Alarm High Red Yes
None,
constant
audible
Message Header: Wrong Power Alarm
Message Details: The unit has detected the
wrong power connected. Verify the proper
voltage.
Ticker Message: Wrong Power Alarm. Press
bell icon for more information.
Wrong Power Alarm
Inactive
None Yellow None None
Message Header: Wrong Power Recovered
Message Details: Wrong Power has been
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Wrong Model Alarm High Red Yes
None,
constant
audible
Message Header: Invalid Control Model
Alarm
Message Details: Invalid Control Model
Alarm. Contact service to ensure the correct
model is selected for the system to avoid
cargo loss.
Ticker Message: Invalid Control Model
Alarm. Press bell icon for more information.
Wrong Model Alarm
Inactive
None Yellow None None
Message Header: Invalid Control Model
Recovered
Message Details: Invalid Control Model has
been recovered.Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Firmware Build
Incompatible
High Red Yes Yes
Message Header: Firmware Build
Incompatible Alarm
Message Details: Firmware build indicates
incompatibility that can result in modules to
be non-coherent.
Ticker Message: Firmware build
incompatible. Press bell icon for more
information.
Alarm Summary
Alarm Message Tone
Heart
Status
Remote
Alarm
Event
Ringback
*
Message Details