EasyManua.ls Logo

Thermo Scientific UltiMate - Probable Causes and Solutions

Thermo Scientific UltiMate
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
Page 44 of 60 M1393R0 OQ/PQ Operating Instructions
5.3 Probable Causes and Solutions
Test Procedure Problem Probable Cause Solution
All
No enough COM ports
available.
- Control Switchos
valves by event
outputs.
Lamp Intensity of the UV
Detector
(4.2)
Intensity values too
low.
Old lamp.
Wrong wavelength.
Integration time too
low.
Replace.
Check/change.
Check/adjust.
Wavelength Check
(4.3)
Failed. Filter not in the
optical path.
Check filter
position.
Flow Cell Check
(4.4)
Intensity values too
low.
See above (4.2).
Dirty flow cell
See above (4.2).
Clean/replace.
UltiMate Fluid Path Test
(4.5)
Flow rate too low. Solvent filter
clogged.
Check/replace.
Baseline Noise and Drift
Test of the UV Detector
(4.6)
Drift too high. Old lamp. Replace.
Oven accuracy
(4.6)
Temperature readout
not within the
specifications.
Temperature probe
not properly
installed.
Install properly (e.g.
use a 4 mm screw).
Linearity of the UV
Detector
(4.7)
Signal is out of range. Acetone
concentration too
high.
Dilute with mobile
phase A.
‘Ghost” peak before
the caffeine peak.
Contaminated
injection valve
Extend wash step.
Sonicate rotor and
stator in CAN for
10 min.
Caffeine peak shows
some ‘spikes’,
improper integration.
Air in the syringe. Check syringe
speed and time
constant (Chapter
6).
Reproducibility of the
Injection Volume
(4.8)
Linearity of the injection
(4.9)
Air in the syringe.
Air at the bottom of
the sample vial.
Purge syringe,
Remove air.
Considerable drift of
the 95% to 100% B
steps.
No/too little formic
acid in mobile
phase A and B.
Check/replace
solvents.
Gradient Accuracy
(4.10)
Result not within
specification
Wrong delay time.
Hardware.
Check/modify.
Check filter, valves,
etc.
Switchos Fluid Path
Check
(4.11)
Flow rate too low. See above (4.5). See above (4.5).
Switchos Flow Rate and
Pressure Stability Test
(4.12)
Pulsation too high. Check valve not
working properly.
Clean/replace.
Switchos Valve Position
Check
(4.13)
Both position LEDs are
illuminated at the
same time.
Improper
initialization or
faulty controller.
See Service
Information #043.

Table of Contents

Related product manuals