Page 46 of 60 M1393R0 OQ/PQ Operating Instructions
6.2 Wavelength Check (NAN Configuration)
Sequence: XQ_1_Wavelength Check_NAN
;Program file for wavelength check
;System must be setup for Nano LC experiments
;Program version 08/10/2004
Sampler.TempCtrl = Off
Oven.TempCtrl = Off
pump.%A.Equate = "100% water + 0.1% FA"
%B.Equate = "Mob. phase A + 0.8% acetone"
%C.Equate = "%C"
%D.Equate = "%D"
Diameter = 75um
Length = 15cm
StationaryPhase = C18_3um_100A
; No automatic CRP calibration!
MasterPressure.LowerLimit = 0.0
MasterPressure.UpperLimit = 400.0
ColumnPressure.LowerLimit = 0.0
ColumnPressure.UpperLimit = 200.0
TrapColumnPressure.LowerLimit = 0.0
TrapColumnPressure.UpperLimit = 200.0
InjectMode = FullLoop
LowDispersionMode = Off
UseAirSegment = Off
UseHeadSpace = Off
SyringeSpeed = Low
SyringeSpeedFactor = 0.1
SampleHeight = 4
FlushVolume = 5.0
WashVolume = 50
RinseBetweenReinjections = Yes
Data_Collection_Rate = 2
pump.Flow = 0.000
%B = 0
%C = 0
%D = 0
;Settings for UV detector
Data_Collection_Rate = 2
UV_VIS_1.Wavelength = 254
TimeConstant = 2.0
0.000 Wait UV.Ready and Pump.Ready and
Pump_FlowSensor.Ready and Sampler.Ready
Inject
Wavelength_Check
UV_VIS_1.AcqOn
1.00 UV_VIS_1.AcqOff
End