EasyManua.ls Logo

Thermo Scientific UltiMate - Performing the UV Detector Linearity Test

Thermo Scientific UltiMate
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test Procedures
Page 32 of 60 M1393R0 OQ/PQ Operating Instructions
4.7.1 Performing the UV Detector Linearity Test
Depending on the system configuration, the ‘UV Detector linearity_NAN’ or ‘UV Detector
linearity_CAP’ sequence is used to perform the detector linearity test. The detector
linearity is determined at 272 nm using caffeine standard samples with different
concentrations:
CAP configuration: 0.25, 0.5, 1.0, 2.0, 5.0, 10.0 and 20.0 µg/mL
NAN configuration: 1.0, 2.0, 5.0, 10.0 , 20.0, 40.0 and 60.0 µg/mL
Install the samples in their proper positions in the FAMOS autosampler rack (TABLE 3-5).
TABLE 2-1 presents the conditions for the linearity test for the two different system
configurations.
TABLE 4-1 Experimental Conditions – UV Detector Linearity
Parameter Nano Configuration Capillary Configuration
Detection wavelength 272 nm 272 nm
Run time 8 min 4 min
Flow rate
0.4 µL/min 4.0 µL/min
CRP 625 50
Injection loop
1.0 µL 1.0 µL
Injection type Partial Loop Fill, 0.2 µL Partial Loop Fill, 0.5 µL
To perform the UV Detector linearity test:
a) Setup the UltiMate system and the FAMOS Microautosampler as discussed in Sections
3.2.2 A and 3.2.2 B.
b) Start the flow delivery of the UltiMate Micropump from the CHROMELEON control panel.
c) After equilibration, start the sequence ‘XQ_3_UV_Linearity_NAN’ (or ‘_CAP’,
respectively).
To evaluate the result of the ‘UV Detector linearity Test’:
a) Use the quantization file ‘caffeine.qnt’.
b) Check the integration and correct if necessary.
c) Verify that the correlation coefficient meets the acceptance criteria (TABLE 2-3).

Table of Contents

Related product manuals