OQ/PQ Operating Instructions M1393R0 Page 47 of 60
6.3 Noise and Drift Test (NAN Configuration)
Sequence: XQ_2 UV Noise and Drift - Oven Test_NAN
Wait Sampler.Ready
;Program file for drift and noise test
;System must be setup for Nano LC experiments
;Program version 08/10/2004
;Settings for restrictor (320cm, 15um ID)
Diameter = 75um
Length = 15cm
StationaryPhase = C18_3um_100A
;CRP = 625
CalibrateCRP When=BeforeFirstSample
;ParkPercentage = Disabled
;Pump limits settings
MasterPressure.LowerLimit = 0.0
MasterPressure.UpperLimit = 400.0
Pump.columnPressure.LowerLimit = 0.0
Pump.columnPressure.UpperLimit = 200.0
TrapColumnPressure.LowerLimit = 0.0
TrapColumnPressure.UpperLimit = 200.0
;Settings for UV detector
Data_Collection_Rate = 2
UV_VIS_1.Wavelength = 254
TimeConstant = 2.0
;FAMOS settings
Wait Sampler.Ready
InjectMode = Fullloop
;Commands for FAMOS with cooling option
;Sampler.TempCtrl = On
;Sampler.Temperature.Nominal = 20.00
;Sampler.Temperature.LowerLimit = 5.00
;Sampler.Temperature.UpperLimit = 30.00
;Virtual channel settings
pump_pressure.Formula Formula=masterpressure.value
pump_pressure.Type = Analog
column_pressure.Formula Formula=pump.columnpressure.value
column_pressure.Type = Analog
;Ultimate pump settings
Pump.%A.Equate = "100% water + 0.1% FA"
Pump.%B.Equate = "Mob. phase A + 0.8% acetone"
Pump.%C.Equate = "%C"
Pump.%D.Equate = "%D"
Pump.Flow = 0.4
%B = 0
%C = 0
%D = 0
0.000 Wait UV.Ready and Pump.Ready and Sampler.Ready
and pump_flowsensor.ready
Inject