Page 56 of 60 M1393R0 OQ/PQ Operating Instructions
%D.Equate = "%D"
0.000 UV.Autozero
pump.Flow = 0.400
%B = 100
%C = 0
%D = 0
Wait UV.Ready and Pump.Ready and
Pump_FlowSensor.Ready and Sampler.Ready
Sampler.Inject
UV_VIS_1.AcqOn
Pump_pressure.AcqOn
Column_pressure.AcqOn
pump.Flow = 0.400
%B = 100
%C = 0
%D = 0
7.000 %B = 100
7.010 %B = 0
30.000 UV_VIS_1.AcqOff
Pump_pressure.AcqOff
Column_pressure.AcqOff
pump.Flow = 0.400
%B = 0
%C = 0
%D = 0
End