Page 14 / 35
INSTRUCTION MANUAL
SAFE 2010/MaxiSAFE 2010
81-09-IM.doc
When the cabinet has been installed:
1. Check that the front window is in the correct position, i.e. parallel to the front shield.
2. Never lift the front window manually, as this will cause the safety latch to engage.
3. Check that the window slides from top to bottom without any irregular sound.
4. If the window does not move when the UP or DOWN button is pressed, authorised service
personnel must be called to correct the fault.
5. Always keep hands and arms away from the working chamber when activating the front
window.
6.3. Connections
Required fusing: Circuit breaker 16 A or fuse T 16 A.
In addition, the applicable safety requirements of the local power Supply Company shall apply.
WARNING
If an automatic relay for disconnection of power in case of electric fault is needed, it
must be installed in the supply, as it is not built-in.
WARNING
Before connecting the power supply it must be checked that the mains
specifications correspond to those stated on the type plate.
WARNING
The safety cabinet is provided with a flexible power cord. The connection may be
installed hard-wired or by means of a wall outlet with protective ground.
If a hard-wired connection is used, a main switch, which will cut off all poles to the unit, must be
used. This switch must be lockable both in the ON and OFF positions.
If the unit is connected through a protective-ground wall outlet, the connector has the all-pole
insulator function. The wall outlet is to be installed out of reach of operators and may only be
accessible to authorised personnel.
7. Testing
Perform tests after installation or change of location and before the initial start-up.
WARNING
According to the standard EN 12469 and DIN 12980, the following must be tested:
• Check that manufacturer’s specifications are met.
• Air velocity in laminar flow.
• Leakage test of the main and exhaust air filter.
• Air inlet speed in the work opening.
• Check airflow patterns.
• Optionally a KI-Discus test for testing operator protection on site.