8.2. While working
WARNING
Important for work with environmentally harmful substances!
• Do not perform work while the fan is running at reduced speed.
• The front window must be in work position during working hours.
• Place the product behind the perforated area of the work surface.
• Work with calm, smooth movements.
• Never overload the work chamber, reduce the number of transfers into and out of the work
chamber.
• Avoid products or agents with high heat emission.
• Avoid a cabinet location where personnel frequently pass and avoid draughts.
WARNING
The efficiency of the laminar airflow in the work chamber is essential for personnel
and product protection. Negative influences of the flow conditions must therefore be
avoided. They primarily arise due to:
• Rapid movements of the operator’s hand, arm, or body both in and in front of the work
chamber.
• Covered vents in the tabletop of the worktable or covered vents at the bottom of the work
chamber back wall.
• Large objects and apparatus.
• Devices making rapid movements e.g. agitators, centrifuges.
WARNING
Do not damage the main filter in the ceiling of the work chamber by mechanical
objects or heat sources, otherwise the microbiological safety is no longer ensured.
The acoustic and optical monitoring devices of the fan and front window must not be deactivated.
The enclosed armrest has to be used in order not to cover the perforated holes.
Devices developing strong heat, e.g. burners. If burners cannot be avoided, use safety burners.
Operate heat sources only with the cabinet fan activated.
8.3. After work
• Remove objects and appliances from the work chamber.
• Clean the work chamber, remove fluids, if any, from the trough and dry it. Disinfect if
required.
• Leave the cabinet fan in operation for about another 10 minutes.
• Close the front window to its lowest position.