EasyManuals Logo

Thermo SAFE 2010 User Manual

Default Icon
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
SAFE 2010/MaxiSAFE 2010
INSTRUCTION MANUAL
Page 9 of 35
81-09-IM.doc
5. Technical specifications
All parts of the cabinet which has a negative pressure is tested in accordance with EN 12469:2000 Annex B,
and meet the LI-C level.
Model 0.9 SAFE MaxiSAFE
External dimensions, DxWxH mm 895x1040x2000
Chamber dimensions, DxWxH mm 580x900x725/650
Weight, kg 200
225
Spillage, litres 13.5
4.5
Exhaust volume, m
3
/h ± 10 %
300
Air velocity in-flow, m/s
0.4
Air velocity down-flow, m/s ± 20 %
0.4
Noise level, DIN dB(A); EN ISO 3744 55
57
Light intensity, Lux >1200
Supply Voltage/frequency, Volts/Hz 230/50
Power consumption, Watts 500
Required fusing Fuse T16A or circuit breaker 16A
Input current, A 1.6
Socket outlet(s), Volts/Hz/Amps 230/50/4
Maximum leakage current, mA 0.7
Model 1.2 SAFE MaxiSAFE
External dimensions, DxWxH mm 895x1340x2000
Chamber dimensions, DxWxH mm 580x1200x725/650
Weight, kg 225
250
Spillage, litres 18
6
Exhaust volume, m
3
/h ± 10 %
400
Air velocity in-flow, m/s
0.4
Air velocity down-flow, m/s ± 20 %
0.4
Noise level, DIN dB(A); EN ISO 3744 57
59
Light intensity, Lux >1200
Supply Voltage/frequency, Volts/Hz 230/50
Power consumption, Watts 600
Required fusing Fuse T16A or circuit breaker 16A
Input current, A 1.8
Socket outlet(s), Volts/Hz/Amps 230/50/4
Maximum leakage current, mA 0.7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo SAFE 2010 and is the answer not in the manual?

Thermo SAFE 2010 Specifications

General IconGeneral
BrandThermo
ModelSAFE 2010
CategorySafety Equipment
LanguageEnglish