SAFE 2010/MaxiSAFE 2010
INSTRUCTION MANUAL
Page 9 of 35
81-09-IM.doc
5. Technical specifications
All parts of the cabinet which has a negative pressure is tested in accordance with EN 12469:2000 Annex B,
and meet the LI-C level.
Model 0.9 SAFE MaxiSAFE
External dimensions, DxWxH mm 895x1040x2000
Chamber dimensions, DxWxH mm 580x900x725/650
Weight, kg 200
225
Spillage, litres 13.5
4.5
Exhaust volume, m
3
/h ± 10 %
300
Air velocity in-flow, m/s
0.4
Air velocity down-flow, m/s ± 20 %
0.4
Noise level, DIN dB(A); EN ISO 3744 55
57
Light intensity, Lux >1200
Supply Voltage/frequency, Volts/Hz 230/50
Power consumption, Watts 500
Required fusing Fuse T16A or circuit breaker 16A
Input current, A 1.6
Socket outlet(s), Volts/Hz/Amps 230/50/4
Maximum leakage current, mA 0.7
Model 1.2 SAFE MaxiSAFE
External dimensions, DxWxH mm 895x1340x2000
Chamber dimensions, DxWxH mm 580x1200x725/650
Weight, kg 225
250
Spillage, litres 18
6
Exhaust volume, m
3
/h ± 10 %
400
Air velocity in-flow, m/s
0.4
Air velocity down-flow, m/s ± 20 %
0.4
Noise level, DIN dB(A); EN ISO 3744 57
59
Light intensity, Lux >1200
Supply Voltage/frequency, Volts/Hz 230/50
Power consumption, Watts 600
Required fusing Fuse T16A or circuit breaker 16A
Input current, A 1.8
Socket outlet(s), Volts/Hz/Amps 230/50/4
Maximum leakage current, mA 0.7