Page 24 / 35
INSTRUCTION MANUAL
SAFE 2010/MaxiSAFE 2010
81-09-IM.doc
11. Let the fan run for 10 minutes.
10.2. Weekly
Wipe the outside of the unit with a mild household cleaner. Anti-static spray may be used to clean
the front window.
10.3. Regularly
Reliable function of the cabinet and compliance with standards are based on the following
conditions:
1. Correct air velocities.
2. Efficiency of the HEPA filters installed.
3. Correctly adjusted alarm limits.
4. Tightness of the construction.
10.4. Decontamination by use of formaldehyde
The purpose of this section is to state guidelines regarding decontamination by using
formaldehyde.
This method is not applicable for cabinets in which activated charcoal filters are installed.
.
NOTE
Certification is required in certain countries for decontamination by using
formaldehyde.
Procedure for decontamination by use of formaldehyde
1. Dismantle the exhaust system.
2. Place the decontamination apparatus on the work surface inside the cabinet.
3. Make sure that the key for the cabinet is in its place.
4. Mount the tent over the cabinet. Be careful so no holes appear in the folio.
• The bottom of the tent is in one piece and two slits must be cut outside of the cabinet
legs, so it is possible to carry the folio to the backside of the cabinet.
• Tape the slits in the bottom folio together with the special tape. Be very careful with the
sealing of the levelling base.
• Pull the wire for the power connection to the front of the cabinet.
• Mount the tent over the cabinet – be careful not to make holes in the folio – and pull it
carefully to the bottom folio.
• Put up the telescopic bars in each corner and tape the tent on to them. Fix the top of the
tent with tape so the tent just reaches the bottom folio.
• Tape the sides of the tent to the bottom folio, starting with the backside of the cabinet,
and take care of a good tightening in the corners.