SAFE 2010/MaxiSAFE 2010
INSTRUCTION MANUAL
Page 31 of 35
81-09-IM.doc
Procedure for replacement of activated charcoal filters
1. Dismantle the cover from the filter box.
2. The charcoal cylinders are fastened with bayonet sockets. Place the special bag over one of
the cartridges. Turn the cartridges clockwise and remove it from the frame. Proceed as
above until all activated charcoal filters have been removed.
3. Install the new charcoal filter cartridges in the filter box.
4. Reinstall the cover.
5. Rinse all tools used. Protective cloth and filter for filter masks are to be placed in a bag for
toxic waste.
11.4. Fuses
The fuses of SAFE 2010/MaxiSAFE 2010 are located in the right side of the top of the cabinet. In
cabinets produced subsequent to August 1999 an additional 2 fuses are located under the
protective cover above the electromechanical panel. These 2 fuses secure zero to light and UV
light. The remaining fuses on the Controller board PCB secure:
Component
Fuse Phase/zero
Main
T10A Phase
Main T10A Zero
Electrical outlet T6.3A Phase
Electrical outlet T6.3A Zero
Option 1 T1A Phase
Option 2 T1A Phase
Option 3 T1A Phase
Light T1A Phase
UV-light T1A Phase
Fan T6.3A Phase
Figure 10a. Figure 10b.
Equipment required