Chapter 3 | Replacing parts
6. Using a pair of pliers, insert the locking pin into the pin locking cylinder arm and
platform locking mechanism and re-position the locking pin (Figure 3.81).
Figure 3.80. Replace pin to locking cylinder
arm and locking mechanism platform.
Figure 3.81. Replacing locking pin to locking
cylinder arm.
3.12. Replacing the 3,000 and 5,000 L assembly
bottom hub clamp
Use the following instructions to replace the assembly bottom hub clamp located on the
bottom of the 3,000 and 5,000 L DynaDrive unit. See Tables 3.1–3.3 for part numbers
based on your system size.
3.12.1. Required parts and tools
• Spare assembly bottom hub clamp (part number SV51245.21)
• 7/32 in. Allen wrench
3.12.2. Preparations
Ensure the BPC is removed from DynaDrive unit before installation.
3.12.3. Replacement procedures
1. Remove the bottom hub clamp assembly by unscrewing the six bolts using the
7/32 in. Allen wrench (Figure 3.82).
Thermo Scientific 46DynaDrive S.U.B. Service Guide