Chapter 3 | Replacing parts
3.11.3. Replacement procedures
1. Using a 3/8 in. Allen wrench, remove the three bolts holding the locking mechanism
platform in place (Figure 3.78).
Figure 3.78. Removing bolts from the
locking mechanism platform.
2. Remove the platform locking mechanism. Replace with new platform locking
mechanism.
3. Using a 3/8 in. Allen wrench, replace the three bolts holding the platform locking
mechanism in place (refer to Figure 3.78).
4. Using a 5/32 in. Allen wrench, replace two bolts to attach the sensor bracket on the
top of the platform locking mechanism (Figure 3.79).
5. Replace pin holding the locking cylinder arm and platform locking mechanism
together (Figure 3.80).
Figure 3.79. Replacing the two bolts to
the sensor bracket.
Thermo Scientific 45DynaDrive S.U.B. Service Guide