Chapter 3 | Replacing parts
2. Using a pair of pliers, locate and remove the locking pin attached to the locking
cylinder arm and platform locking mechanism (Figure 3.74).
Figure 3.74. Removing locking pin from
locking cylinder arm.
3. Remove pin holding the locking cylinder and locking mechanism platform together.
Push the locking cylinder to the side (Figure 3.75).
Figure 3.75. Removing pin from locking
cylinder arm and locking mechanism
platform.
4. Using a 5/32 in. Allen wrench, remove two bolts to detach the sensor bracket from
the top of the locking mechanism platform (Figures 3.76 and 3.77).
Figure 3.76. Removing the two bolts
from the sensor bracket.
Figure 3.77. Detaching the sensor
bracket.
Thermo Scientific 44DynaDrive S.U.B. Service Guide